Clinical Result of Excimer Laser Photoastigmatic Refractive Keratectomy Using VISX 20/20 B VisionKeyTM with Version 4.01 Software
- Affiliations
-
- 1Department of Ophthalmology, College of Medicine, Yeungnam University, Taegu, Korea.
Abstract
-
Photoastigmatic refractive keratectomy (PARK) using VISX 20/20 B excimer laser with version 4.01 software was performed in 205 consecutive eyes of compound myopic astigmatism to assess the efficacy, predictability, and safety of excimer laser treatment of compound myopic astigmatism between September 1994 and June 1996. The mean preoperative manifest spherical equivalent refraction was -6.95+/-2.52D (range -1.13 to -15D) and the mean preoperative manifest cylindrical refraction was -1.25+/-0.76D (range -0.5 to -4.50D). The patients were followed from 6 months to 1 year (mean 9.19 months). At 1 year, 82(75.2%) of all 109 eyes receiving PARK were within 1D of plano refraction, and 88(80.7%) of all 109 eyes achieved uncorrected visual acuity of 20/30 or better. Postoperative refractions were stable after 3 months without significant early overcorrection. Mean postoperative astigmatism was -0.29+/-0.40D at 6 months and -0.24+/-0.42D at 12 months. Overall improvement of astigmatism was 75% at 6 months and 82% at 12 months by vector analysis using Alpins` method. In conclusion, PARK using VISX 20/20 B VisionKeyTM excimer laser with version 4.01 software appears to be effective in the treatment of compound myopic astigmatism with a relatively high degree of accuracy and safety. The predictability and stability of the postoperative refraction during the first 12 months seem to be quite reliable.