J Korean Acad Oral Maxillofac Radiol.
1999 Feb;29(1):161-173.
A Comparative Study of Quantitative Assessment of Bone Mineral Density of Mandible
- Affiliations
-
- 1Department of Oral and Maxillofacial Radiology, College of Dentistry, Chosun University, Korea.
Abstract
-
This study was performed to compare the bone mineral densities measured at mandibular premolar area by copper-equivalent image and hydroxyapatite phantom with those measured at radius by dual energy absorptiometry and to evaluate the clinical usefulness of Digital system with slide scanner, copper-equivalent image, and hydroxyapatite phantom.
For experiment, intraoral radiograms of 15 normal subjects ranged from 20 years old to 67 old were taken with copper-step wedge at mandibular premolar area and bone mineral densities calculated by conversion equation to bone mineral density of hydroxyapatite were compared with those measured at radius distal 1/3 area by Hologic QDR-1000.
Obtained results as follows:
1) The conversion equation was Y=5.97X-0.25 and its determination coefficient was 0.9967. The coefficient of variation in the measurement of copper-equivalent value ranged from 4% to 8% and showed high reproducibility.
2) The coefficient of variation in the measurement of bone mineral density by the equation ranged from 7% to 8% and showed high reproducibility.
3) The bone mineral densities ranged from 0.35 to 0.79g/cm2 at mandibular premolar area.
4) The correlation coefficient between bone mineral densities at mandibular premolar area and those at radius distal 1/3 area was 0.8965.
As summary, digital image analyzing system using copper-equivalent image and hydroxyapatite phantom appeared to be clinically useful to measure the bone mineral density at dental area.