J Korean Acad Prosthodont.  2002 Oct;40(5):470-483.

Corrosion Phenomena Of Magnetic Attachment Used In Oral Environment

Affiliations
  • 1Department of Prosthodontics and Oral Biology Research Institute, College of Dentistry, Chosun University. jhajung@mail.chosun.ac.kr
  • 2Department of Dental Laboratory Technology and Metallogical Engineering, Kwangyang University.

Abstract

The purpose of this study was to investigate further the mechanics of failure of magnets used for denture retention. Dyna magnets were retrived from denture that had failed after 34 months of clinical use. The magnets were observed and sectioned in order to analyse with high resolution scanning electron microscope. From this study, corrosion behaviors of used magnetic attachment were analysed. The results were as follows: 1. In Nd-Fe-B based magnetic materials, the erosion-corrosion was started at ununiformed part of stainless steel cover. 2. Corrosion was initiated at weared stainless steel surface and then magnetic materials were spalled by corrosive solution. 3. Spalling was occurred in Nd-Fe-B magnet materials due to corrosion products and then corrosion rate was increased drastically. 4. Corrosion started from ununiformed stainless steel surface as well as welded zone. In conclusion, the failure of magnets may occur by either breakdown of the welding or breakdown of the encapsulating material. So, it is considered that the corrosion problem of dental magnetic materials could be solved to some extent with surface treatment of dental magnetic materials.

Keyword

Magnetic attachment; Dyna magnets; Corrosion behaviors; Spalling

MeSH Terms

Corrosion*
Denture Retention
Dentures
Mechanics
Stainless Steel
Welding
Stainless Steel
Full Text Links
  • JKAP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr