Korean Circ J.  1987 Mar;17(1):103-112. 10.4070/kcj.1987.17.1.103.

Comparison of Two-Dimensional Doppler Echocardiography and Fick Method with Cardiac Catheterization of Qp, Qs and Qp/Qs Ratio in Ventricular Septal Defect in Children

Abstract

This study was performed to compare the amount of blood flow from left and right side heart using Doppler echocardiography and Fick method. The pulmonary blood flow (Qp), aortic blood flow (Qs) and Qp/Qs ratio were noninvasively evaluated by two dimensional Doppler echocardiography in 10 patients with ventricular septal defect in children, aged 2 years to 15 years (mean+/-SD=7.1+/-4.9 years). Cardiac catheterization was performed within 2 days after examination of Doppler echocardiography in all patients. Systemic and pulmonary blood flow was calculated by using Fick principle, based on measured oxygen consumption and measured blood oxygen saturation. The Doppler frequency shift was analyzed by computer system combined with Meridian Echocardiography System. Comparison of Qp, Qs and Qp/Qs ratio obtained by invasive methods and by two-dimensional pulsed wave Doppler echocardiography showed a correlation of 0.679 in Qp, 0.710 in Qs and 0.639 in Qp/Qs ratio (p<0.05).

Keyword

Doppler echocardiography; Aortic blood flow; Pulmonary blood flow; Ventricular septal defect; Cardiac catheterization

MeSH Terms

Cardiac Catheterization*
Cardiac Catheters*
Child*
Computer Systems
Echocardiography
Echocardiography, Doppler*
Heart
Heart Septal Defects, Ventricular*
Humans
Oxygen
Oxygen Consumption
Oxygen
Full Text Links
  • KCJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr