Korean J Anesthesiol.  2007 Nov;53(5):676-679. 10.4097/kjae.2007.53.5.676.

Fires and Burn Caused by Alcohol-based Disinfectants and Electrocautery: A case report

Affiliations
  • 1Department of Anesthesia and Pain Medicine, Holy Family Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea. eychung@hfh.cuk.ac.kr

Abstract

Operating room fires, though rare, can involve substantial morbidity and mortality. Surgical fires require an ignition source, oxidizer, and fuel. Ignition sources generally include lasers and electrocautery, oxidizer are usually oxygen, nitrous oxide, and ambient air, whereas fuels are classically surgical drapes, materials, and prepping agents. We experienced a patient who, during skin incision, sustained burns resulting from a fire in the operating room. Shortly after application of disinfectants and placement of the surgical drapes, the surgeon used the electrosurgical unit on the incision. In this case, the use of an alcohol-based disinfectant was the major contributing factor to the surgical fire. To avoid recurrence, if alcohol is used for skin prepping, it should be allowed to dry completely before draping.

Keyword

alcohol; electrosurgical unit; fire

MeSH Terms

Burns*
Disinfectants*
Electrocoagulation*
Fires*
Humans
Mortality
Nitrous Oxide
Operating Rooms
Oxygen
Recurrence
Skin
Surgical Drapes
Disinfectants
Nitrous Oxide