Korean Circ J.  1997 May;27(5):501-507. 10.4070/kcj.1997.27.5.501.

Evaluation of Correlation between Automatic Ocillometric Sphygmomanometer and Standard Korotkoff Auscultatory Sphygmomanometer

Abstract

BACKGROUND
Early diagnosis and treatment of hypertension is imperative to prevent the complications associated with this condition. The development of accurate and convenient methods of blood pressure measurement, therefore, is indispensible. At present, the JNC V has acknowledged the use of automatic sphygmomanometer that can be used without the help of someone else. We compared automatic oscillometric sphygmomanometers manufactured by Sein Electronics, korea(SE-7000 and SE-5000) with the standard Korotkoff auscultatory mercuy sphygmomanometer meter. The correlation between these two methods were calculated to determine whether these products could actually be used in clinical practice. This study was undertaken to ensure the clinical evaluation of these two products and standardization of an antomatic sphygmomanometer in korea before it is actually used.
METHODS
The study included eighty-three patients, ranging in age from 14 to 81 years, who were admitted to Hanyang University Hospital in October, 1995. The blood pressure measured by the automatic oscillometric sphygmomanometer(SE-7000) in the right arm and was compared with that measured by the standard mercury sphygmomanometer(baumanometer) in the left arm. Also the blood pressure measured by the automatic oscillometric sphygmomanometer(SE-5000) at the wrist was compared by the baumanometer in the same side arm. The correlation between these methods were determined by the paired Student`s t-test and by the simple liner regression method.
RESULTS
The p value of systolic blood pressure between two methods(SE-7000 and baumanometer) in the both arms was 0.896 and correlation coefficient was 0.8286. The p value of diastolic blood pressure between this two methods was 0.352 and correleation coefficient was 0.7455. The p value of systolic blood pressure between two methods(SE-5000 and baumanomter) by the arm and the wrist was 0.00018 and correlation coefficient was 0.8588, the p value of diastolic blood pressure between this two methods was 0.000048 and correlation coefficient was 0.5944.
CONCLUSION
The blood pressure measured by the SE-5000 at the wrist was statistically different from that measured with the baumanometer in the arm. Further studies are necessary to use this product in clinical practice. However, the systolic and diastolic blood pressures measured with the SE-7000 in the arm were relatively similar to those measured by the baumanometer enabling this products to be effectively used in clinical practice.

Keyword

Oscillometric sphygmomanometer; Korotkoff auscultatory sphygmomanometer

MeSH Terms

Arm
Blood Pressure
Early Diagnosis
Humans
Hypertension
Korea
Sphygmomanometers*
Wrist
Full Text Links
  • KCJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr