J Korean Radiol Soc.  2002 Jul;47(1):35-42. 10.3348/jkrs.2002.47.1.35.

Stent Insertion in Patients with Malignant Biliary Obstruction: Problems of the Hanaro Stent

Affiliations
  • 1Department of Radiology, Kyungpook National University School of Medicine, Korea. sck@knu.ac.kr
  • 2Department of Radiology, Kosin National University School of Medicine, Korea.
  • 3Department of Radiology, Dong-A National University School of Medicine, Korea.

Abstract

PURPOSE: To investigate the problems of the Hanaro stent (Solco Intermed, Seoul, Korea) when used in the palliative treatment of patients with inoperable malignant biliary obstruction.
MATERIALS AND METHODS
Between January 2000 and May 2001, the treatment of 46 patients with malignant biliary obstruction involved percutaneous placement of the Hanaro stent. Five patients encountered problems during removal of the stent's introduction system. The causes of obstruction were pancreatic carcinoma (n=2), cholangiocarcinoma (n=2), and gastric carcinoma with biliary invasion (n=1). In one patient, percutaneous transhepatic cholangiography and stent insertion were performed as a one-step procedure, while the others underwent conventional percutaneous transhepatic biliary drainage for at least two days prior to stent insertion. A self-expandable Hanaro stent, 8-10 mm in deameter and 50-100 mm in lengh, and made from a strand of nitinol wire, was used in all cases.
RESULTS
Among the five patients who encountered problems, breakage of the olive tip occourred in three, upward displacement of the stent in two, and improper expansion of the distal portion of the stent, unrelated with the obstruction site, in one. The broken olive tip was pushed to the duodenum in two cases and to the peripheral intrahepatic duct in one. Where the stent migrated during withdrawal of its introduction system, an additional stent was inserted. In one case, the migrated stent was positioned near the liver capsule and the drainage catheter could not be removed.
CONCLUSION
Although the number of patients in this study was limited, some difficulties were encountered in withdrawing the stent's introduction system. To prevent the occurrence of this unusual complication, the stent should be appropriately expansile, and shape in the olive tip shoud be considered.


MeSH Terms

Catheters
Cholangiocarcinoma
Cholangiography
Drainage
Duodenum
Humans
Liver
Olea
Palliative Care
Seoul
Stents*
Full Text Links
  • JKRS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr