Yonsei Med J.  2013 Sep;54(5):1289-1292. 10.3349/ymj.2013.54.5.1289.

Spurious Elevation of Glucose Concentration during Administration of High Dose of Ascorbic Acid in a Patient with Type 2 Diabetes on Hemodialysis

Affiliations
  • 1Department of Internal Medicine, Gyeongsang National University School of Medicine, Jinju, Korea. jrhahm@gnu.ac.kr
  • 2Institute of Health Sciences, Gyeongsang National University School of Medicine, Jinju, Korea.
  • 3Department of Laboratory Medicine, Gyeongsang National University School of Medicine, Jinju, Korea.
  • 4Department of Biochemistry, Gyeongsang National University School of Medicine, Jinju, Korea.

Abstract

We describe herein a case of life-threatening hypoglycemia due to spurious elevation of glucose concentration during the administration of ascorbic acid in a type 2 diabetic patient. A 31-year-old female was admitted for proliferative diabetic retinopathy treatment and prescribed high dose ascorbic acid. During hospitalization, she suddenly lost her consciousness and her glucose concentration was 291 mg/dL, measured using self-monitoring blood glucose (SMBG) device, while venous blood glucose concentration was 12 mg/dL. After intravenous injection of 50% glucose solution, the patient became alert. We reasoned that glucose measurement by SMBG device was interfered by ascorbic acid. Physicians should be aware of this interference; high dose ascorbic acid may cause spurious elevation of glucose concentration when measuring with SMBG devices.

Keyword

Self-monitoring of blood glucose; ascorbic acid; hypoglycemia

MeSH Terms

Adult
Ascorbic Acid/administration & dosage/adverse effects/contraindications/*therapeutic use
Blood Glucose
Blood Glucose Self-Monitoring/instrumentation/standards
Diabetes Mellitus, Type 2/blood/drug therapy
Female
Humans
Hypoglycemia/*diagnosis
Renal Dialysis
Ascorbic Acid
Blood Glucose

Cited by  1 articles

Influence of Vitamin C and Maltose on the Accuracy of Three Models of Glucose Meters
Jooyoung Cho, Sunyoung Ahn, Jisook Yim, Younjung Cheon, Seok Hoon Jeong, Sang-Guk Lee, Jeong-Ho Kim
Ann Lab Med. 2016;36(3):271-274.    doi: 10.3343/alm.2016.36.3.271.


Reference

1. Martin S, Schneider B, Heinemann L, Lodwig V, Kurth HJ, Kolb H, et al. Self-monitoring of blood glucose in type 2 diabetes and long-term outcome: an epidemiological cohort study. Diabetologia. 2006; 49:271–278.
Article
2. Welschen LM, Bloemendal E, Nijpels G, Dekker JM, Heine RJ, Stalman WA, et al. Self-monitoring of blood glucose in patients with type 2 diabetes who are not using insulin: a systematic review. Diabetes Care. 2005; 28:1510–1517.
Article
3. Davis WA, Bruce DG, Davis TM. Is self-monitoring of blood glucose appropriate for all type 2 diabetic patients? The Fremantle Diabetes Study. Diabetes Care. 2006; 29:1764–1770.
Article
4. Melker RJ. Test strips for blood glucose monitors are not always accurate. Diabetes Care. 2003; 26:3190.
Article
5. Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol. 2009; 3:903–913.
Article
6. Arabadjief D, Nichols JH. Assessing glucose meter accuracy. Curr Med Res Opin. 2006; 22:2167–2174.
Article
7. Tang Z, Du X, Louie RF, Kost GJ. Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer. Am J Clin Pathol. 2000; 113:75–86.
Article
8. Heller A, Feldman B. Electrochemical glucose sensors and their applications in diabetes management. Chem Rev. 2008; 108:2482–2505.
Article
9. Frias JP, Lim CG, Ellison JM, Montandon CM. Review of adverse events associated with false glucose readings measured by GDH-PQQ-based glucose test strips in the presence of interfering sugars. Diabetes Care. 2010; 33:728–729.
Article
10. Melamud B, Shilo S, Munter G. Life-threatening hypoglycemia due to false measurement of glucose in a peritoneal dialysis patient. Isr Med Assoc J. 2010; 12:125–126.
11. International Organization for Standardization. In vitro diagnostic test systems-Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switserland: International Organization for Standardiation;2003.
12. Riley SG, Chess J, Donovan KL, Williams JD. Spurious hyperglycaemia and icodextrin in peritoneal dialysis fluid. BMJ. 2003; 327:608–609.
Article
13. Disse E, Thivolet C. Hypoglycemic coma in a diabetic patient on peritoneal dialysis due to interference of icodextrin metabolites with capillary blood glucose measurements. Diabetes Care. 2004; 27:2279.
Article
14. Yoo EH, Lee SY. Glucose biosensors: an overview of use in clinical practice. Sensors (Basel). 2010; 10:4558–4576.
Article
15. Lyon ME, DuBois JA, Fick GH, Lyon AW. Estimates of total analytical error in consumer and hospital glucose meters contributed by hematocrit, maltose, and ascorbate. J Diabetes Sci Technol. 2010; 4:1479–1494.
Article
Full Text Links
  • YMJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr