Korean J Med Phys.  2005 Jun;16(2):89-96.

An Effect of Time Gating Threshold (TGT) on the Delivered Dose at Internal Organ with Movement due to Respiration

Affiliations
  • 1Department of Biomedical Engineering, School of Medicine, The Catholic University of Korea, Korea. suhsanta@catholic.ac.kr
  • 2Department of Radiation Sciences, Seoul Health College, Korea.
  • 3Department of Radiation Oncology, Konkuk University College of Medicine, Korea.

Abstract

In this study, we investigated the effect of time gating threshold on the delivered dose at a organ with internal motion by respiration. Generally, the internal organs have minimum motion at exhalation during normal breathing. Therefore to compare the dose distribution time gating threshold, in this paper, was determined as the moving region of target during 1 sec at the initial position of exhalation. The irradiated fields were then delivered under three conditions; 1) non-moving target 2) existence of the moving target in the region of threshold (1sec), 3) existence of the moving target region out of threshold (1.4 sec, 2 sec). And each of conditions was described by the moving phantom system. It was compared with the dose distributions of three conditions using film dosimetry. Although the treatment time increased when the dose distributions was obtained by the internal motion to consider the TGT, it could be obtained more exact dose distribution than in the treatment field that didn't consider the internal motion. And it could be reduced the unnecessary dose at the penumbra region. When we set up 1.4 sec of threshold, to reduce the treatment time, it could not be obtained less effective dose distribution than 1 sec of threshold. Namely, although the treatment time reduce, the much dose was distributed out of the treatment region. Actually when it is treated the moving organ, it would rather measure internal motion and external motion of the moving organ than mathematical method. If it could be analyzed the correlation of the internal and external motion, the treatment scores would be improved.

Keyword

Internal organ motion; Time gating threshold (TGT); Respiration

MeSH Terms

Exhalation
Film Dosimetry
Respiration*
Full Text Links
  • KJMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr