Korean J Med Phys.  2005 Jun;16(2):77-81.

Characteristics of Detectors for Measurements of Photon Depth Doses in Build-Up Region

Affiliations
  • 1Department of Radiation Oncology, College of Medicine, Hallym University, Korea. bychul@hallym.or.kr

Abstract

To determine the appropriate method out of various available methods to measure build-up doses, the measurements and comparisons of depth doses of build-up region including the surface dose were executed using the Attix parallel-plate ionization chamber, the Markus chamber, a cylindrical ionization chamber, and a diode detector. Based on the measurements using the Attix chamber, discrepancies of the Markus chamber were within 2% for the open field and increased up to 3.9% in the case of photon beam containing the contaminant electrons. The measurements of an cylindrical ionization chamber and a diode detector accord with those of the Attix chamber within 1.5% and 1.0% after those detectors were completely immersed in the water phantom. The results suggest that the parallel-plate chamber is the best choice to measure depth doses in the build-up region containing the surface, however, using cylindrical ionization chamber or diode detector would be a reasonable choice if no special care is necessary for the exact surface dose.

Keyword

Build-up; Electron contamination; Ionization chamber; Diode detector