Korean J Nucl Med.
2000 Jun;34(3):222-227.
Left Ventricular Ejection Fraetion Determmed by Cated Tl-201
Perfusion SPECT and Quantitative Software
Abstract
-
PURPOSE: We compared estimates of ejection fraction (EF) determined by gated Tl-201
perfusion SPECT (g-TI-SPECT) with those by gated blood pool (GBP) scan.
MATERIALS AND METHODS
Eighteen subjects underwent g-TI-SPECT and GBP scan.
After reconstruction of g-TI-SPECT, we measured EF with Cedars software. The comparison
of the EF with g-TI-SPECT and GHP scan was assessed by correlation analysis and Bland
Altman plot.
RESULTS
The estimates of EF were significantly different (p<0.05) with g-TI-SPECT
(40%+/-14%) and GBP scan (43%+/-14%). There was an excellent correlation of EF between
e-TI-SPECT and GBP scan (r=0.94, p<0.001). The mean difference of EF between GRP scan
and g-TI- SPECT was +3.2%, Ninety-five percent limits of agreement were +9,8%. EF between
g-TI-SPECT and GBP scan were in poor agreement.
CONCLUSION
The estimates of EF by g-TI-SPECT was well correlated with those by GBP scan.
However, EF of g-TI-SPECT doesn't agree with EF of GBP scan. EF of g-TI-SPECT cant be used
interchangeably with EF of GBP scan.