J Korean Soc Magn Reson Med.  2007 Jun;11(1):39-48.

The Comparison between Single Shot Turbo Spin Echo and B-FFE (Balanced Turbo Field-echo) in the Differentiation of Focal Liver Lesions

Affiliations
  • 1Department of Diagnostic Radiology, Yonsei University College of Medicine, Korea. radpms@yumc.yonsei.ac.kr
  • 2Department of Diagnostic Radiology, Yonsei University Wonju College of Medicine, Korea.
  • 3Department of Internal Medicine, Yonsei University College of Medicine, Korea.
  • 4Department of General Surgery, Yonsei University College of Medicine, Korea.

Abstract

PURPOSE: To determine the diagnostic accuracy of four different sequences : moderately T2 weighted, two heavily T2-weighted single shot turbo spin-echo sequence and breath-hold axial-2D balanced turbo field-echo sequence(bFFE) for characterization of focal lesions.
MATERIALS AND METHODS
During the 3-month period between June and August 2005, seventy-six patients were proved to have ninety-three focal hepatic lesions on MR imaging. The patients consisted of 49 men and 27 women (age range, 15-75 years; mean age, 56.23 years). All MR images were acquired on a 1.5-T MR using the following sequences: 1. A breath-hold axial T2-weighted single shot turbo spin-echo sequence, 2. a breath-hold axial-2D balanced turbo field-echo sequence. Two radiologists performed quantitative analysis. Another radiologist measured the lesion-to-liver contrast-to-noise ratio at the region-of-interest in the four sequences.
RESULTS
There was no significant difference in inter-observer variability between the four sequences. The accuracy for both cyst and malignancy of moderate T2 weighted MRI (echo time: 80 msec) was also highest. There was significant difference for lesion characterization between moderate T2 weighted MRI and balanced steady state procession (p-value: 0.004) in the second reader. For longer echo time, the CNR of cystic lesions were markedly increased in comparison to lesions of other component.
CONCLUSION
The accuracy and inter-observer variability of single shot turbo spin echo T2 weighted sequence was higher than bFFE. Although there was no statically significant difference, moderate T2 weighted MRI (echo time: 80 msec) was more accurate than heavily T2 weighted sequence (echo time: 300 msec). If the results for lesion characterization is equivocal in TE 80, the addition of heavily T2 weighted MRI (echo time: 180 msec) can be helpful.

Keyword

Heavily T-2 weighted MRI, turbo spin echo; Steady state free procession, echo time

MeSH Terms

Female
Humans
Liver*
Magnetic Resonance Imaging
Male
Observer Variation
Full Text Links
  • JKSMRM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr