J Korean Acad Prosthodont.  2008 Apr;46(2):209-216.

The influence of abutment screw tightening timing and DLC coating of conical connection implant system

Affiliations
  • 1Department of Prosthodontics, Graduate School, Seoul National University, Seoul, Korea. 0504heo@hanmail.net

Abstract

STATEMENT OF PROBLEM: It is difficult about assessing the loosening tendency of conical connection type implant after cyclic loading with diamond like carbon coating one-piece abutments, and also about the retightening effect. PURPOSE: This study was performed to investigate the influence of one-piece abutment screw retightening after 5.0X10(4) cyclic loading and Diamond like coating MATERIAL AND METHODS: Thirty two ITI implant were divided to 4 groups. Group 1,3-titanium abutment, group 2,4 - diamond like carbon coated abutment. Group 1,2 - 20.0X10(4) cyclic loading after 5.0X10(4) cyclic loading, Group 3,4- after 20.0X10(4) Cyclic loading. After cyclic loading, periotest values were taken and removal torque values of abutments were measured with a digital torque gauge.
RESULTS
1. The removal torque of group 2 after 5.0X10(4) cyclic loading is slightly greater than the other groups but not significantly higher than others (P>0.05). 2. The final removal torque values after 20.0X10(4) cyclic loading of group 1 is bigger than group 3, and group 2 is bigger than group 4, but not significantly higher (P>0.05). 3. The final removal torque values after 20.0X10(4) cyclic loading of all groups are not significantly different (P>0.05).

Keyword

Conical connection; One-piece abutment; Diamond-like coating; Cyclic loading; Periotest

MeSH Terms

Carbon
Diamond
Torque
Carbon
Diamond