J Korean Neurol Assoc.  2014 Feb;32(1):8-13.

Feasibility of 1.6-MHz Probe for Detection of Cerebral Blood Flow in the Poor Temporal Window

Affiliations
  • 1Department of Neurology, Ajou University School of Medicine, Suwon, Korea. dacda@hanmail.net

Abstract

BACKGROUND
The demographics of the stroke population consist mainly of elderly patients. Transcranial Doppler is an effective method for intracranial flow detection, but can be limited due to the poor temporal window (PTW), which is frequent in the elderly. Therefore, we investigated whether the low frequency 1.6-MHz probe can be useful to improve flow detection for PTW.
METHODS
All 201 participants had a history of transient ischemic attack or minor ischemic stroke. Firstly, we analyzed the success rate of recording intracranial blood flow via temporal window by using 2.0-MHz (high frequency, HF) and 1.6-MHz (low frequency, LF) probes for a maximum of 10 minutes. Secondly, mean flow velocity (MFV) and pulsatile index (PI) of insonated spectrum of 25 patients with good temporal window were compared between the two probes.
RESULTS
Seventy-eight patients were female and the mean age was 64.1+/-12.4 years. Among 402 windows, 125 (31.1%) were undetected when using the 2.0-MHz probe. Fifty-five patients were detected for the spectrum by the 1.6-MHz probe. However, 70 (17.4%) still remained as PTW. Between the two probes, there were no significant differences of variables from the middle cerebral arteries: MFVs (HF 61.0+/-14.1 vs. LF 61.3+/-14.8 cm/sec, p=0.403 in the right; HF 59.6+/-13.4 vs. LF 59.3+/-13.3 cm/sec, p=0.232 in the left) and PIs (HF 0.82+/-0.17 vs. LF 0.82+/-0.18, p=0.929 in the right; HF 0.82+/-0.20 vs. LF 0.83+/-0.17, p=0.605 in the left).
CONCLUSIONS
Conventional 2.0-MHz probe had relatively high proportion of PTW but 1.6-MHz probe was feasible to improve flow detection for PTW.

Keyword

Cerebral arteries; Temporal bone; Transcranial doppler sonography

MeSH Terms

Aged
Cerebral Arteries
Demography
Female
Humans
Ischemic Attack, Transient
Middle Cerebral Artery
Stroke
Temporal Bone
Ultrasonography, Doppler, Transcranial
Full Text Links
  • JKNA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr